Mispa Revo Plus

Mispa Revo Plus

Time-Resolved Fluorescence Immunoassay (TRFIA) Analyzer.

Overview

Introducing the Mispa Revo Plus – a state-of-the-art Time Resolved Fluorescence Immunoassay (TRFIA) analyzer. Designed to provide precise analysis for a wide range of parameters, including Thyroid, fertility markers, cardiac marker, Iron and Born Metabolism Tumor Marker, Inflammation, and more

Accurate Results Anytime, Anywhere: Mispa Revo Plus

The Mispa Revo Plus employs Time-Resolved Fluorescence Immunoassay (TRFIA) technology, enhancing sensitivity and specificity through the use of rare earth ion fluorescence. This super-sensitive technique overcomes the instability of enzyme markers, making it ideal for immunological detection.
TRFIA’s high fluorescence intensity and prolonged signal duration reduce interference from fluorescent substances, ensuring accurate results.

With 18-month cartridge stability, Mispa Revo Plus may test more than 20 parameters in various profiles, including those for the thyroid, heart, fertility, bone metabolism (vitamin D), and others. With a time constraint of fewer than 20 minutes, the rapid and secure QR Code calibration aids in getting the report.

Technical Specifications

Method

Time-Resolved Fluorescence Immunoassay (TRFIA)

Specimen

Serum/Plasma/ Whole Blood

Weight

1.3 kg

Dimensions

225mmx152mmx105mm (LxWxH)

Screen

7 inch touch screen

Data Storage

10000 Nos.

Printer

Built-in thermal printer

Battery

Built-in lithium battery (standby time: lOhrs)

Power Source

AClO0V – 240V AC, S0-60Hz, 60VA

Communication

Rs232 (LIS/HIS), USB

 

Features:

  • TRFIA Technology
  • Compact and Portable
  • 18 Parameters Assay Panel
  • Better sensitivity and specificity.
  • Secure QR Code Calibration
  • 6 Hours of battery back-up
  • 7inch LCD Touch Screen Display
  • LIS/HIS Available
  • Stable Test cards with 24 Months of stability
  • Can be performed in Serum and Plasma
  • Easy Operation
  • Built in Thermal Printer
  • 10000 patient result storage
  • Testing  & Validated with CLIA method.

 

Sl.No.

Category

Parameter

Measuring range

Sample vol. (µl)

Incubation (min)

Sample type 

1

Thyroid

TT3

0.5-10 nmol/L

100 µl

15 min

serum/plasma

2

TT4

10-350 nmol/L

100 µl

10 min

serum/plasma

3

TSH

0.1-60 µIU/ml

100 µl

15 min

serum/plasma

4

Cardiac

NT-proBNP

50-25000 pg/ml

100 µl

15 min

serum/plasma

5

D-Dimer

0.1-10 µg/ml

100 µl

10 min

Plasma

6

Trop I

0-40 ng/ml

100 µl

10 min

serum/plasma

7

Fertility Hormones

LH

1-200 mIU/ml

100 µl

15 min

serum 

8

FSH

0.6-200 mIU/ml

100 µl

15 min

serum

9

AMH

0.1-50 ng/ml

100 µl

10 min

serum

10

PRL

25-4000 µIU/ml

100 µl

15 min

serum

11

Prog

0.3-50 ng/ml

100 µl

10 min

serum/plasma

12

β- HCG

2-20000 mIU/ml

100 µl

15 min

serum/plasma

13

Inflammatory markers

IL-6

4.8-2000 pg/ml

100 µl

15 min

serum

14

CRP

0.5-200 µg/ml

5 µl

3 min

serum/plasma/WB

15

PCT

0.1-50 ng/ml

100 µl

10 min

serum/plasma

16

Calcium Metabolism

Vit-D

5-70 ng/ml

5 µl

10 min

serum/plasma/WB

17

Iron Metabolism

Ferritin

10-1000 ng/ml

5 µl

15 min

serum/plasma

18

Cancer Marker

PSA

0.1-100 ng/ml

100 µl

10 min

serum/plasma

Test Menus – Pack Size – 25 Tests

Technology:

There are two main types of immunofluorescence assays: direct and indirect. Direct FIA uses a single antibody but is more expensive and less sensitive. Indirect FIA, which uses two antibodies, is more cost-effective and offers higher sensitivity.
The Mispa Revo Plus utilizes the Time-Resolved Fluorescence Immunoassay (TRFIA) technique, an advanced form of indirect FIA, providing superior sensitivity and specificity.
Our cartridges feature rare earth ions (Eu3+) as the fluorescent compound. These ions produce a highly specific and prolonged fluorescent signal, ensuring accurate and reliable results.